Devices

A 14 × 14 multiplexed surface array with 196 platinum sites on 250 µm center pitch, 29 interface wires and a roughly 25 µm flexible film. Rat auditory-cortex recordings, 2014.

Device — Cortical surface

Active micro-ECoG array (196 sites, auditory cortex)

micro-ECoG · multiplexed · active electronics · auditory cortex · Connecticut · Pennsylvania · NYU · academic
3D GEOMETRY · REFERENCE MODEL

Active micro-ECoG · 196-site recording patch

Electrode-bearing structure · dimensions in millimeters

Loading 3D viewer…

Physical sites / interface wires196 / 29; active multiplexing
Layout14 × 14 sites, 250 µm center pitch
Site size200 × 200 µm
Recording-area envelopeApproximately 3.5 × 3.5 mm
Multilayer thicknessApproximately 25 µm
Omitted hardwareSwitching layers, full film outline and cable

Model fidelity. Flat recording-patch reference for the 2014 auditory-cortex array. Site count, matrix, pitch, size and multilayer thickness come from the paper. The 3.5 mm square uses the reported approximate recording-area envelope as a crop, not a recovered full film outline. Platinum contact faces are shown as surfaces without adding thickness above the published multilayer total. Switching electronics, interconnects, flexible cable and curved cortical placement are omitted. Electrical channel timing and tissue response are not modeled.

Sources & reference images

Dimensions come from the linked specifications; images guide shape only. Reference images remain on their owners’ websites.

JSON includes physical site coordinates, unknown channel mappings and approximation notes. GLB uses meters. These models do not predict recording or stimulation performance.

Active micro-ECoG array for auditory-cortex recordings

A flexible surface array with switching electronics at each recording site. Multiplexing reduces the external wiring: this 196-site device uses 29 interface wires instead of a separate wire for every site.

Published geometry

Field2014 paper
Site arrangement14 × 14 matrix, 196 electrodes
Site size200 × 200 µm
Gap between sites50 µm
Center-to-center pitch250 µm
Recording areaApproximately 3.5 × 3.5 mm
Multilayer thicknessApproximately 25 µm
Interface wires29
Recording surfacePlatinum, approximately 50 nm deposited layer
Site impedanceApproximately 45 kΩ at 1 kHz

The paper distinguishes this active array from the passive NeuroNexus E32-300-20-50 used in a comparison group. That passive array has 32 sites and is not the hardware described here.

What was shown

The team used the custom active array in three anesthetized rats and the passive array in three other rats. The active array mapped sound responses across auditory cortex. Frequency-response maps were compared with intrinsic optical imaging from the same cortical locations.

This is an acute experiment, not chronic implant evidence. The paper’s discussion of scaling to thousands of sites is a design argument, not a demonstrated thousand-site implant.

Place in surface-array history

This device follows the actively multiplexed flexible arrays described by Viventi and colleagues in 2010 and 2011. The 2014 paper uses smaller pitch and a different 196-site layout. It belongs in Devices because it specifies a distinct hardware configuration, even though auditory mapping is the experiment used to validate it.

Organizations

The paper lists University of Connecticut, New York University, University of Pennsylvania, University of Illinois at Urbana-Champaign and Northwestern University among its US affiliations, alongside collaborators in South Korea and China. These are paper affiliations, not claims about who manufactured a commercial product.

Sources and model limits

  • Escabí MA, Read HL, Viventi J, et al. A high-density, high-channel count, multiplexed μECoG array for auditory-cortex recordings. Journal of Neurophysiology 112:1566-1583 (2014). DOI: 10.1152/jn.00179.2013. Primary manuscript, fabrication and recording methods.

A recording-patch reference model can preserve the published site geometry and film thickness. The full flexible outline, interface cable and switching layers need separate reconstruction; they should not be guessed from the recording-area size.